cropped cropped 小 lo副本 副本
YOKY

Fully Automatic Closed Cup Flash Point Tester BBS-800

Fully Automatic Closed Cup Flash Point Tester BBS-800

Fully Automatic Closed Cup Flash Point Tester BBS-800

This fully automatic closed flash point tester is manufactured in accordance with the new standard method 1 of GB/T261-2021 “Determination of Flash Point by Binsky Martin Closed Cup Method” and the requirements of ASTMD93. It is used in departments such as petroleum, chemical, power, environmental protection, railway, and scientific research, and is an ideal instrument for conducting closed flash point testing.

1. Adopting a new high-speed signal processor control, adaptive PID control algorithm, automatically adjusting the heating curve according to standard requirements;
2. Equipped with an automatic atmospheric pressure detection chip, atmospheric pressure is automatically measured, calibrated, and the flash point value is automatically corrected;
4. Safe without gas source, platinum alloy electric heating wire ignition method, safe, stable, reliable, high testing accuracy;
5. Differential detection, automatic correction of system deviation, automatic stop detection and alarm for temperature exceeding the limit;;
6. The interface can be operated in both Chinese and English, and the instrument program can be upgraded during the revision of national standards;
7. Optional USB interface can export data to a computer (optional), RS485 port can communicate with the user
8. Comes with 12 sets of data for temperature point correction
9. The furnace body temperature is automatically displayed, and the furnace body temperature can be checked at any time for easy sampling next time.
10. Meet the new GB/T261-2021 standards 261-A, 261-B, 261-C, and optional testing per degree.
Step A: The sample is heated at a rate of 5 ℃/min to -6 ℃/min, with a stirring speed of 90r/min to 120 r/min
Step B: Heat the sample at a rate of 1 ℃/min to 1.5 ℃/min, stirring at a speed of 250r/min ± 10 r/min
Step C: The sample is heated at a rate of 2.5 ℃/min to 3.5 ℃/min, with a stirring speed of 90r/min to 120 r/min

 

Model BBS-800
Display 8-inch 65K (resolution 800 * 600) touch screen
Measurement method thermocouple differential detection method
Display range -59.9~400℃
Measurement range RT -400 ℃ (to determine the flash point at minus zero temperature, the sample needs to be frozen first)
Resolution 0.1 ℃
Repeatability Repeatability: “National standard: allowable deviation of ± 1 ℃ when flash point value<104 ℃, allowable deviation of ± 2 ℃ when flash point value ≥ 104 ℃
International standard: When the flash point value is ≤ 110 ℃, the allowable deviation is ± 1 ℃, and when the flash point value is>110 ℃, the allowable deviation is ± 2 ℃“
Heating rate in accordance with GB/T261-2021, step A, step B, step C, test per degree
Printer low energy high-speed thermal printer, 56mm paper width
Temperature detection imported platinum resistor (Pt100 Ω)
Ignition method no gas source , electronic ignition
Self checking function automatic diagnosis of instrument faults
Cooling method forced air cooling
Dimensions 450X280X290mm
Weight 15kgs